Model PMD-60E
6 Digit
Product Description
Panel mount, 1/8 DIN standard, larger cutout for optional metal case, for pressure/load sensors, 0.5 to 20 mV/V strain gage, optional voltage DC or AC, Current.
Performance Specifications
DESCRIPTION: | LED Digital Display, 7-segment, 0.56”, 6 digits |
RECOMMENDED INSTALLATION: | Panel mount, 1/8 DIN standard, larger cutout for optional metal case |
HOUSING MATERIAL: | 94V-0 UL rated self-extinguishing polycarbonate, optional metal case |
APPROVALS: | CE per EN-61000-3/4/6 and EN-61010-1 |
OPERATING TERMPERATURE: | 0° to 50°C (32 °F to 122 °F) |
STORAGE TEMPERATURE: | -20° to 70°C (-4°F to 15 8°F) |
RELATIVE HUMIDITY: | 95% (non-condensing) at 40°C (104°F) |
ENVIROMENTAL: | Optional IP 65 ore NEMA-4 front cover |
WEIGHT: | 0.79 lbs (11 .5 oz) standard, weight increases with options |
OVERALL DIMENSIONS: | 3.78”W x 1.89”H x 5.61” D (96x48x143 mm) including rear connectors |
TRANSDUCER ACCEPTED: | Pressure/load sensors, 4-wire full bridge or 6-wire with internal shunt |
TRANSDUCER INPUTS ACCEPTED: | 0.5 to 20 mV/V strain gage, optional Voltage DC or AC, Current |
TRANSDUCER EXITATION: | Jumper selectable 5 VDC, 10 VDC or 24 VDC |
OPTIONAL SENSORS ACCEPTED: | RTD, Thermocouple, LVDT, potentiometric, magnetostrictive, frequency, pH |
ANALOG OUTPUT: | 0 to 5/10 VDC or 4-20 mA, or option for both (output does not go negative) |
ANALOG OUTPUT RESOLUTION: | 16-bit D/A provides 25 0μV for voltage output or 0.4μA for current output |
ANALOG OUTPUT RESPONSE: | Typical 143 msec, faster response optional |
PEAK CAPTURE RESPONSE: | Typical 330 msec, faster response optional up to 10 msec |
COMMUNICATIONS: | Optional RS232, RS485, Ethernet, DeviceNet, ModBus |
DIGITAL DISPLAY: | Red LED standard, green or super bright optional, numeric or alpha |
DISPLAY UPDATE RATE: | Typical 5 per second |
DISPLAY POLARITY INDICATION: | Assumed positive, displays (-) or negative |
OVERRAGE INDICATION: | Displays (-------) |
DECIMAL POINT SELECTION: | Front panel, user selectbale to five positions |
ZERO ADJUSTMENT: | Digital/manual entry or apply zero input to sensor |
SPAN ADUJUSTMENT: | Digital/manual entry or apply full scale input to sensor |
SCALING ADJUSTMENT: | Digital/manual entry or apply full scale input to sensor |
POWER SUPPLY: | 85 to 265 VAC, optional 10 to 72 VDC |
SHUNT CALIBRATION: | Remote shunt or optional shunt resistor internal to meter |
OTHER FEATURES: | RS232 programmable, Peak/Valley, Linearization, Math Functions, Data Logging |
OTHER OPTIONS: | Multible displays, Dual/Tri/Quad inputs, up to 6 limits with programmable setpoints,PID loop control, optically isolated I/O, custom macro programming |